1 / 3
Linshang
Authorized brand
LS183 spectrum transmission meter employing the parallel light path source. is mainly used for the optical performance test of thermal insulation coatings, single-layer glass, laminated glass (insulating glass), glass with film, PMMA, PC and other materials. It can simultaneously test the transmittance of ultraviolet, infrared and visible light. The width of the test slot is 47mm. The meter uses ultraviolet light source, infrared light source and visible light source to illuminate the tested transparent material. The sensor determines the transmittance by detecting the original light intensity emitted by the three light sources and the transmitted light intensity after the light has passed through the transparent material. The final result is expressed as a percentage.
UV Peak Wavelength 365nm
IR Peak Wavelength 940nm
Visible Light Peak Wavelength 380nm-760nm (conform to CIE photopic luminosity function)
Resolution 0.1%
Measurement Accuracy ± 2%(colorless and uniform transparent materials)
Test Sample Thickness < 47mm
Dimension 216 mm * 134 mm * 29 mm (L*W*H)
Test Slot Dimension 47mm * 91 mm(W*H)
Weight About 590g
Power Supply DC5V/1A
DC5V Adapter,User manual, Support pillar